Lambert Beer Law

 

The Absorbance of a solution

For each wavelength of light passing through the spectrometer, the intensity of the light passing through the reference cell is measured. This is usually referred to as Io – that’s I for Intensity.

The intensity of the light passing through the sample cell is also measured for that wavelength – given the symbol, I. If I is less than Io, then obviously the sample has absorbed some of the light. A simple bit of math is then done in the computer to convert this into something called the absorbance of the sample – given the symbol, A.

For reasons to do with the form of the Beer-Lambert Law (discussed below), the relationship between A (the absorbance) and the two intensities is given by:

A=log(IoI)

On most of the diagrams you will come across, the absorbance ranges from 0 to 1, but it can go higher than that.

An absorbance of 0 at some wavelength means that no light of that particular wavelength has been absorbed. The intensities of the sample and reference beam are both the same, so the ratio Io/I is 1. Log10 of 1 is zero. An absorbance of 1 happens when 90% of the light at that wavelength has been absorbed – which means that the intensity is 10% of what it would otherwise be.

In that case:

IoI=10010=10

and

log1010=1.

 

The importance of concentration

The proportion of the light absorbed will depend on how many molecules it interacts with. Suppose you have got a strongly colored organic dye. If it is in a reasonably concentrated solution, it will have a very high absorbance because there are lots of molecules to interact with the light.

However, in an incredibly dilute solution, it may be very difficult to see that it is colored at all. The absorbance is going to be very low. Suppose then that you wanted to compare this dye with a different compound. Unless you took care to make allowance for the concentration, you couldn’t make any sensible comparisons about which one absorbed the most light.

 

The importance of the container shape

Suppose this time that you had a very dilute solution of the dye in a cube-shaped container so that the light traveled 1 cm through it. The absorbance isn’t likely to be very high. On the other hand, suppose you passed the light through a tube 100 cm long containing the same solution. More light would be absorbed because it interacts with more molecules.

Again, if you want to draw sensible comparisons between solutions, you have to allow for the length of the solution the light is passing through. Both concentration and solution length are allowed for in the Beer-Lambert Law.

 

The Lambert Beer Law

You will find that various different symbols are given for some of the terms in the equation – particularly for the concentration and the solution length. I’m going to use the obvious form where the concentration of the solution is “c” and the length is “l”.

Lambert Beer Law

You should recognize the expression on the left of this equation as what we have just defined as the absorbance, A. You might also find the equation written in terms of A:

A=ϵlc

That’s obviously easier to remember than the first one, but you would still have to learn the equation for absorbance. It might be useful to learn it in the form:

A=log(IoI)=ϵlc 

The Greek letter epsilon in these equations is called the molar absorptivity – or sometimes the molar absorption coefficient.

 

Molar absorptivity

If you rearrange the simplest of the equations above to give an expression for epsilon (the molar absorptivity), you get:

ϵ=Alc

Remember that the absorbance of a solution will vary as the concentration or the size of the container varies. Molar absorptivity compensates for this by dividing by both the concentration and the length of the solution that the light passes through. Essentially, it works out a value for what the absorbance would be under a standard set of conditions – the light travelling 1 cm through a solution of 1 mol dm-3.

That means that you can then make comparisons between one compound and another without having to worry about the concentration or solution length.

Values for molar absorptivity can vary hugely. For example, ethanal has two absorption peaks in its UV-visible spectrum – both in the ultra-violet. One of these corresponds to an electron being promoted from a lone pair on the oxygen into a pi anti-bonding orbital; the other from a pi bonding orbital into a pi anti-bonding orbital.

The table gives values for the molar absorptivity of a solution of ethanal in hexane. Notice that there are no units given for absorptivity. That’s quite common. If you want them, and assuming the length is in cm and the concentration is mol dm-3, the units are mol-1 dm3 cm-1.

electron jump wavelength of maximum absorption (nm) molar absorptivity
lone pair to pi anti-bonding orbital 290 15
pi bonding to pi anti-bonding orbital 180 10000

The ethanal obviously absorbs much more strongly at 180 nm than it does at 290 nm. (Although, in fact, the 180 nm absorption peak is outside the range of most spectrometers.)

You may come across diagrams of absorption spectra plotting absorptivity on the vertical axis rather than absorbance. However, if you look at the figures above and the scales that are going to be involved, you aren’t really going to be able to spot the absorption at 290 nm. It will be a tiny little peak compared to the one at 180 nm. To get around this, you may also come across diagrams in which the vertical axis is plotted as log10(molar absorptivity).

If you take the logs of the two numbers in the table, 15 becomes 1.18, while 10000 becomes 4. That makes it possible to plot both values easily, but produces strangely squashed-looking spectra